Preparation
Full and even deflation to ensure no folds obstruct the view of the camera
Generous water based lubricant to the posterior aspect of the cuff only to avoid blurring the camera view
Videoscope placed into the 'video' channel
Ensure availability of TRI-1/TRI-2 connector
Operation
Insert slowly and visualize landmarks
After passing the epiglottis, adjust the camera angle for full view of the glottis
Inflate cuff according the recommended volume and check pressure with a cuff manometer
Removal
Press release buttons on both sides of the scope to release the videoscope from the mask
Remove SGA using standard technique